Page 68 - Doedijns_Temperature_Guide_Book_LR_pag_los

This is a SEO version of Doedijns_Temperature_Guide_Book_LR_pag_los. Click here to view full version

« Previous Page Table of Contents Next Page »
TEMPERATURE SENSORS
8. A FEW CONSTRUCTIONS OF
68
Type ‘A’
Type ‘A’ Mi-cable (MgO) thermocouples are free hanging for use
in non pressurized applications.
Type ‘B’
Type ’B’ Mi-cable (MgO) thermocouples can be retracted for
transport and mounting purposes. Individually replaceable
without disassembling the complete unit.
Type ‘C’
Type ‘C’ Mi-cable (MgO) thermocouples are spring loaded to con-
nect with the inner wall of the thermowell. This construction
allows replacement of the thermocouple during operation.
Type ‘D’
Each Type ‘D’ Mi-cable (MgO) or beaded type thermo- couple is
enclosed in a guide tube which terminates at a special block,
welded to the thermowell at the point to be measured.
Each thermocouple can be replaced during operation.
MULTI POINT